Valve module for compressed air-driven sensor cleaning in the IQ SENSOR NET, System 28X


Design according to the minimum requirements defined by ATV (Waste water association) 
in bulletin M256 for meters to be used at waste water treatment plants.

Dual IQ / valve module for compressed air-driven sensor cleaning in the System 28X.
The valve module is time-controlled by a relay of the DIQ/S 28X measuring transmitter.

The cleaning is done by means of external compressed air; the compressed air stream is effective 
via a separate cleaning head that is screwed on the sensor or directly at the UV/VIS sensors.

The external compressed air supply can be provided via a compressed air line or a compressor.
Requirements for this: dry compressed air, free of oil and dust, 5 bar, approx. 40 NL/min

Power supply:
Directly via the auxiliary voltage of the DIQ/S 28X measuring transmitter

Instrument protection:
IP 66 type of protection

Enclosure material:
Polystyrol

Dimensions:
94 x 94 x 57 mm (W x H x D)

Manufacturer:
WTW

Type:
DIQ/CHV
or equivalent

			